Welcome to Dexsa Utama Indonesia

KYORITSU 4102A-H

• 3-Pole and 2-Pole Earth Resistance
• The latest circuit design permits the instrument to operate with the minimum of influence from earth voltage and earth resistance of auxiliary earth spikes
• Dust and drip proof (designed to IEC 60529 IP54)
• Earth resistance value can be read directly from the scale
• Designed to meet IEC 61010-1 safety standard
• Capable of measuring earth voltage
• Small and lightweight. Shock resistant new case material
• 2mA measuring current permits earth resistance tests without tripping earth leakage current breakers in the circuit under test
• Lead wire connection to C and P terminals and proper auxiliary earth resistance can be checked by “OK” lamp
Lead wire connection to C and E terminals is good when “OK” lamp is illuminated

Berat1 kg

Specifications

Earth resistance
measurement
× 1ΩRange× 10Ω× 100Ω
Measuring range0 – 12Ω0 – 120Ω0 – 1200Ω
Accuracy±3% of full scale
Earth voltage measurement
Measuring range0 – 30 V AC (50/60Hz)
Accuracy±3% of full scale
Overload protectionEarth resistance : 276V AC/DC (10 seconds)
Earth voltage : 276V AC/DC (10 seconds)
Applicable StandardsIEC 61010-1 CAT Ⅲ300 V Pollution degree 2
IEC 61010-2-030, IEC 61557-1,5, IEC 60529 (IP54)
Power sourceR6(AA)(1.5V) × 6
Dimension105(L) × 158(W) × 70(D) mm (including case lid)
WeightApprox. 600g (including batteries and case lid)
Accessories7095A (Earth resistance test leads (Red 20m, Yellow 10m, Green 5m/1set))
7127B (Simplified measurement probe)
8032 (Auxiliary earth spikes[2spikes/1set])
9121(shoulder strap)
9084 (Carrying case)
R6 (AA) × 6, Instruction manual
Optional Accessories7245A (Precision measurement cord set)
8259 (Adapter for measurement terminal)
9164 (Carring case [Hard])

Brosure

KYORITSU 4102A

Ulasan

Belum ada ulasan.

Hanya pelanggan yang sudah login dan telah membeli produk ini yang dapat memberikan ulasan.

SKU: KYORITSU 4102A-H Kategori: Tag: , ,

Click here to assign a widget for this area.